Study summary
| Patient ID |
KHIS001 |
Patient name |
(hidden — destination not de-identified)
|
| Study UID |
1.2.840.113654.2.4.4.3.4.119950730134200 |
Study date |
19950730 |
| Description |
OFFIS Structured Reporting Samples |
Accession number |
KHIS073013420 |
| Content |
2 series, 2 instance(s), modalities:
CR, SR |
Failed transfers |
0 |
Content sent
|
SOP Class |
UID |
| Basic Text SR Storage |
1.2.840.10008.5.1.4.1.1.88.11 |
| Computed Radiography Image Storage |
1.2.840.10008.5.1.4.1.1.1 |
|
Transfer syntax |
UID |
| Implicit VR Little Endian |
1.2.840.10008.1.2 |
| Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
Series
|
Series UID |
Modality |
SOP Class(es) |
Instances |
|
1.2.276.0.7230010.3.1.3.1787205428.166.1117461927.66 |
SR |
1.2.840.10008.5.1.4.1.1.88.11 |
1 |
|
1.2.840.113654.2.4.4.3.5.119950730134200 |
CR |
1.2.840.10008.5.1.4.1.1.1 |
1 |
Conformance findings per SOP Class
Basic Text SR Storage
1.2.840.10008.5.1.4.1.1.88.11
|
Severity |
Issue |
Tag |
Attribute / Module |
Expected |
Found |
Instances |
|
|
|
(0070,0022) |
Graphic Data |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0070,0023) |
Graphic Type |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0040,A168) |
Concept Code Sequence |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0040,A130) |
Temporal Range Type |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A130) |
Temporal Range Type |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A168) |
Concept Code Sequence |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0040,A801) |
Tabulated Values Sequence |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0040,A050) |
Continuity Of Content |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (3006,0024) |
Referenced Frame of Reference UID |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0008,1199) |
Referenced SOP Sequence |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0008,1199) |
Referenced SOP Sequence |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0070,0023) |
Graphic Type |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0070,0022) |
Graphic Data |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(3006,0024) |
Referenced Frame of Reference UID |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A801) |
Tabulated Values Sequence |
Type 1: present with a value |
Attribute is missing |
1 |
|
|
|
(0040,A730) > (0040,A300) |
Measured Value Sequence |
Type 2: present, possibly empty |
Attribute is missing |
1 |
|
|
|
(0040,A300) |
Measured Value Sequence |
Type 2: present, possibly empty |
Attribute is missing |
1 |
Computed Radiography Image Storage
1.2.840.10008.5.1.4.1.1.1
|
Severity |
Issue |
Tag |
Attribute / Module |
Expected |
Found |
Instances |
|
|
|
(0028,3000) > (0028,3006) |
LUT Data |
Type 1: present with a value |
Attribute is present but empty |
1 |
|
|
|
(0018,1152) |
Exposure |
an integer in -2^31..2^31-1 |
2.500000 |
1 |
|
|
|
(0020,0060) |
Laterality |
a Laterality for the paired body part PED CHEST |
no laterality is given |
1 |
|
|
|
(0028,0200) |
Image Location |
Attribute is retired in the current DICOM standard |
Attribute is used |
1 |
|
|
|
(0018,1010) |
Secondary Capture Device ID |
an attribute defined in the IOD of this SOP Class |
standard attribute not part of this IOD (Standard Extended SOP Class) |
1 |
|
|
|
(0018,1210) |
ConvolutionKernel |
an attribute defined in the IOD of this SOP Class |
standard attribute not part of this IOD (Standard Extended SOP Class) |
1 |
Study consistency
No consistency issue across the dataset.
How to read this report
|
|
Violation of the DICOM standard: missing/empty Type 1 attribute, missing mandatory module, wrong VR or VM. |
|
|
Recommended-practice deviation: missing Type 2 attribute, unexpected enumerated value, consistency issue. An
empty Type 2 attribute is conformant. |
|
|
Informational only: retired attributes, unknown VR. |
| Issue colors |
|
| Notes |
Conditional modules and Type 1C/2C attributes are validated only when present (their conditions are free text in the standard). De-identification may legitimately empty or remove attributes; all values shown are the de-identified
ones actually sent. Pixel data and other bulk values are not inspected; private tags are not validated.
|
|