Karnak DICOM Conformance Report
Destination Local DICOM Folder: ./dicom  •  Source AET test
FAILED
  3 error(s)   6 warning(s)   4 info
Generated 2026-07-04 08:54:09 CEST  •  Karnak Development  •  innolitics/dicom-standard JSON (vendored 2025-02)

Study summary

Patient ID PIG 4BA_98659 Patient name (hidden — destination not de-identified)
Study UID 1.2.840.113619.2.98.3467.1098086125.0.69 Study date 20041018
Description Test Photometric Interpretation Accession number
Content 6 series, 6 instance(s), modalities: US Failed transfers 0

Content sent

SOP Class UID
Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1
Ultrasound Multi-frame Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.3
Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1
 
Transfer syntax UID
Explicit VR Little Endian 1.2.840.10008.1.2.1
JPEG Baseline (Process 1) 1.2.840.10008.1.2.4.50

Series

Series UID Modality SOP Class(es) Instances
1.2.40.0.13.1.1.45653013396044006507995806643500112598 US 1.2.840.10008.5.1.4.1.1.3.1 1
1.2.40.0.13.1.1.122270019448124560098735282914374064195 US 1.2.840.10008.5.1.4.1.1.3 1
1.2.40.0.13.1.1.174837738883920071074308671435667278685 US 1.2.840.10008.5.1.4.1.1.6.1 1
1.2.840.113619.2.98.3467.1098086125.0.70 US 1.2.840.10008.5.1.4.1.1.6.1 1
1.2.40.0.13.1.1.289494807713901808938101707293962434443 US 1.2.840.10008.5.1.4.1.1.3.1 1
1.2.40.0.13.1.1.292543359504713203703828512757607614864 US 1.2.840.10008.5.1.4.1.1.6.1 1

Conformance findings per SOP Class

Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1

Severity Issue Tag Attribute / Module Expected Found Instances
ERROR
Missing module
cine Mandatory module (usage M) None of its attributes is present. Required attributes: Multiplexed Audio Channels Description Code Sequence 1
ERROR
Pixel geometry
(0028,0034) Pixel Aspect Ratio absent when the ratio is 1:1 (the implied default) 1\1 1
WARNING
Private creator missing
(140D,0010) Private Creator A Private Creator reserving this block No Private Creator defines this private block 1
WARNING
Private creator missing
(160D,0010) Private Creator A Private Creator reserving this block No Private Creator defines this private block 1
WARNING
Private creator missing
(170D,0010) Private Creator A Private Creator reserving this block No Private Creator defines this private block 1
WARNING
Type 2 missing
(0008,0008) Image Type Type 2: present, possibly empty Attribute is missing 1

Ultrasound Multi-frame Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.3

Severity Issue Tag Attribute / Module Expected Found Instances
WARNING
Retired SOP Class
(0008,0016) SOP Class UID A non-retired SOP Class Ultrasound Multi-frame Image Storage (Retired) is retired: IOD module checks skipped 1
WARNING
Value format
(0008,0023) Content Date a valid date YYYYMMDD 1994.11.05 1
INFO
Retired attribute
(0008,2204) Transducer Orientation Attribute is retired in the current DICOM standard Attribute is used 1

Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1

Severity Issue Tag Attribute / Module Expected Found Instances
ERROR
Pixel geometry
(0028,0034) Pixel Aspect Ratio absent when the ratio is 1:1 (the implied default) 1\1 1
INFO
Retired attribute
(0010,1000) Other Patient IDs Attribute is retired in the current DICOM standard Attribute is used 1
INFO
Non-standard attribute
(0018,1004) Plate ID an attribute defined in the IOD of this SOP Class standard attribute not part of this IOD (Standard Extended SOP Class) 1
INFO
Non-standard attribute
(0040,1400) Requested Procedure Comments an attribute defined in the IOD of this SOP Class standard attribute not part of this IOD (Standard Extended SOP Class) 1

Study consistency

No consistency issue across the dataset.

How to read this report

ERROR
Violation of the DICOM standard: missing/empty Type 1 attribute, missing mandatory module, wrong VR or VM.
WARNING
Recommended-practice deviation: missing Type 2 attribute, unexpected enumerated value, consistency issue. An empty Type 2 attribute is conformant.
INFO
Informational only: retired attributes, unknown VR.
Issue colors
presence
encoding (VR/VM)
value constraints
consistency
retired usage
Notes Conditional modules and Type 1C/2C attributes are validated only when present (their conditions are free text in the standard). De-identification may legitimately empty or remove attributes; all values shown are the de-identified ones actually sent. Pixel data and other bulk values are not inspected; private tags are not validated.
Copyright Karnak Team